Datasheet PR515x504

Signal converters and isolating amplifiers

Issued 3 November 2026

Thermocouple to 4-20 mA, galvanically isolated, with OLED display

Article number PR515x504

PR515 module board from above: OLED display, UP, OK and DWN buttons, relay, USB-C port and terminals 1 to 22

The PR515x504 measures the temperature with a thermocouple and passes it on as 4-20 mA. The input is galvanically isolated from the supply and the output; this is needed with a thermocouple that touches metal at the measuring point, such as a welded or grounded type. The thermocouple type (K, J, T, N, E, R, S or B) and the temperature at 4 and at 20 mA are set on the OLED display.

Price on request. Supplied in series, from 10 units per item. Not in stock? Then we build your series. One or a few units only as a trial, for a company with a series in sight.

Specifications

Supply12 and 24 V DC, operating range 11 to 30 V DC
InputThermocouple type K, J, T, N, E, R, S or B, type selected on the display; with cold-junction compensation
Measuring rangeAdjustable within the range of the selected type, for example 0 to 600 °C
Output4-20 mA, active
IsolationThermocouple input galvanically isolated from supply and output
SetupOLED display with UP, DOWN and OK buttons
USBUSB-C for software updates and reading the status
Wi-FiBuilt-in hotspot for software updates
MountingDIN rail, horizontal enclosure
Internally fitted withPR181T51

Setup

  1. Connect the supply, the thermocouple and the output according to the wiring diagram supplied with the module.
  2. On the display, select the thermocouple type.
  3. Set the temperature at 4 mA and at 20 mA and store with SAVE.